Pilot Point of Care C-Reactive Protein (CRP) Testing EQA

Accreditation Status: currently not accredited to ISO 17043
Date Scheme started: 2017
Clinical Applicability: Monitoring of the acute phase response
Analytes: C-Reactive Protein (POC/555)
Units for Reporting: mg/L
Samples Distributed:

Liquid format. Normal and pathological human serum.

Additional materials may be produced for specific recovery experiments by the addition of purified CRP to an analyte-free serum matrix.

Number of Distributions per year: 4
Number of Samples per Distribution: 2
Frequency of Distributions: Currently seasonal as outlined in the Distribution Schedule
Schedule of Analysis:

Data Entry is via the web for the submission of results. Data analysis commences 21 days after sample dispatch. Late returns are accepted and will contribute to the laboratory's cumulative performance statistics

Data Analysis:

All Laboratory Trimmed Mean (ALTM) with truncation at 2SD, 3SD, and CV%.

Reports also show method specific statistics. Individual laboratory performance is expressed in terms of MRBIS, SDBIS, and MRVIS

Chosen Coefficient of Variation for C-Reactive protein is 8%

Performance Scoring: MRVIS
Criteria of Performance:

Laboratory performance is classified in terms of the MRVIS over a running analytical window of 4 Distributions (12 months)

  MRVIS
Ideal <50
Good 50-100
Adequate 101-200
Poor >200 or SDBIS > 200

 

Persistent Poor Performance: